Accessories

5.8 Air/Water-Cooled Housing

The Air/Water-Cooled Housing accessory allows the sensor to be used in ambient temperatures up to
120°C (250°F) with air-cooling, and 180°C (356°F) with water-cooling. The cooling media should be
connected using 1/8" NPT fittings. Air flow should be 1.4 to 2.5 l/sec at 25°C (77°F). Water flow should
be approximately 1.0 to 2.0 l/min (water temperature between 10 and 27°C (50 to 80.6°F). Chilled water
below 10°C (50°F) is not recommended, see section 5.8.1 Connecting, page 40.
The Air/Water-Cooled Housing is available in stainless steel (...S) and aluminum (...A) and includes the
air purge collar.
Sensor with DIN quick connector

5.8.1 Connecting

As a standard the cooled housing is supplied with fittings and non-metric 1/8" NPT threads. If you
intend to supply the cooling media through pipes or hoses, consider the following information.
Thread Adapter - metric pipe to NPT inside thread
Thread Adapter - inch-type pipe to NPT inside thread
Supporting tube for PVC or Tygon hoses
40
Figure 34: Air/Water-Cooled Housing
Rev. B2 Feb 2017
Sensor with terminal strip connection
Inner Thread NPT (in.)
1/8
1/8
1/8
Inner Thread NPT (in.)
1/8
Tube Outer Ø
6 mm (0.24 in)
8 mm (0.31 in)
mm [in]
Pipe Outer Ø
4 mm (0.16 in)
6 mm (0.24 in)
8 mm (0.31 in)
DIN-ISO-Outer Thread
4 mm (0.16 in)
Tube Inner Ø
4 mm (0.16 in)
6 mm (0.24 in)
XR