When using the Right Angle Mirror, adjust the emissivity settings downward by 5%.
For example, for an object with an emissivity of 0.65, you adjust the value down to 0.62.
This correction accounts for energy losses in the mirror.

5.5 Sighting Viewer

The Sighting Viewer (XXXTXXACSV) is used to aid in the alignment of the standard sensor. It is used
when the object is small and far from the sensor as well as when it is difficult to obtain a direct in-line
sighting. It can be used both with and without the Air Purge Collar, but not with the Right Angle Mirror.
First secure the sensor to the bracket using the mounting nut or the Air Purge Collar. Next screw on the
Sighting Viewer Tool fully. Now, position and secure the bracket. When the alignment is complete, do
not forget to remove the Sighting Viewer Tool.
The Sighting Viewer Tool has the same dimensions as the Right Angle Mirror.

5.6 Adjustable Pipe Adapter

The adjustable pipe adapter (XXXTXXACPA) is available to connect a sighting tube or conduit to a
sensor with or without water/air cooled housing.
38
Figure 32: Right Angle Mirror
Figure 33: Adjustable Pipe Adapter