VOLTAGE CHECK: The tester first checks for possible interference. If there
is an external voltage of more than 42V, the warning triangle
displayed. If measurements are being performed in AUTO mode and a
disturbance frequency is detected, the tester will automatically look for a
frequency that is different from the default 128Hz.

5.1 Switching the Test Voltage

Available in all functions except mΩ and 2 Clamp.
If necessary, the test voltages can be switched to either 16 or 32V as follows:
• Press the Hz/OPTIONS button, then press it again to make the output volt-
age (Uout) blink.
• Use the ► button to switch to 16 or 32V, then press Hz/OPTIONS again.
This setting is retained when the unit is turned off.
5.2 Resistance Measurement mΩ (2-Wire)
WARNING: Before performing the resistance test, verify that the sample
under test is not energized.
1.
Set the switch to
2.
Connect the resistance to terminals H (Z) and E (X).