If you need to change the Object number, press the ▲ button until the desired
number is displayed. Then press the ► button; the Test number now blinks.
•
Press the ▲ button until the desired test is displayed.
Press the MEM button to display the dEL screen. By default, the blinking word
•
NO appears on the screen, indicating the test is not to be deleted. To delete
the test, press the ▲ button to change the blinking NO to YES, then press and
hold down (>2s) the MEM button. To exit without erasing, perform a short press
on the MEM button.
Note that if the test stored in this Object/Test number is a sweep test (which con-
sists of multiple measurements taken at different frequencies) this procedure only
deletes a single measurement. The remaining sweep measurements will still be
stored in this Object/Test. You must delete all individual measurements to com-
pletely remove a sweep test.
When a test is deleted, its Object/Test number is removed from memory. When you
subsequently view stored tests, this number will be skipped. This Object/Test num-
ber combination remains unavailable for storing tests until you completely erase
memory, as instructed in Section 7.3.1.