AEMC 6471 User Manual page 46

Digital ground resistance and soil resistivity tester
Hide thumbs Also See for 6471:
Table of Contents

Advertisement

Observe the following minimum distances between the two clamps in order to
avoid direct electromagnetic influences between the transmitting clamp and the
receiving clamp:
Value measured
0 to 1
1 to 5
5 to 10
10 to 50
50 to 100
100 to 500
1.
Start the measurement by pressing the START/STOP button.
In the case of the diagram below, the earth impedance measured is equal to:
R
= R
+ ( R
LOOP
E
NOTE: In the AUTO mode, the measurement frequency is 1611Hz. To make
an earth measurement that is free of inductive effects, you must change to
MANUAL mode and choose a lower measurement frequency (see §6.2).