Setting Parameters For Fid - Agilent Technologies 7820A Advanced User's Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

5
Detectors
* Detector hydrogen is pressure controlled, where the detector provides a known resistance. If using a nickel catalyst tube, the resistance
changes, and the flow rates displayed by the GC will not be accurate. Measure the actual hydrogen flow using a flow meter at the detector
vent, with all other flows turned off.

Setting parameters for FID

WA RN IN G
70
Use nitrogen makeup gas (instead of helium) for greater
sensitivity.
Verify that a column is installed or the FID column fitting is
plugged before turning on the air or hydrogen. An explosion may
occur if air and hydrogen are allowed to leak into the oven.
To set the FID parameters:
1
Verify:
Makeup gas is configured
Installed jet type is correct for column type
Press [Front Det] or [Back Det].
2
3
Set the detector temperature. The temperature must be
greater than 150 °C for the flame to light.
Set the hydrogen flow rate, if desired, and press [Off/No].
4
Change the air flow rate, if desired, and press [Off/No].
5
If using a packed column, set the FID makeup gas to 0.0/Off.
6
If using a defined capillary column, set the makeup gas flow
7
or combined column plus makeup gas flow.
Scroll to Flame and press [On/Yes]. This turns on the air and
8
hydrogen and initiates the ignition sequence. The signal
typically increases to 5 to 20 pA after ignition.
Verify that the flame is lit by holding a cold, shiny surface,
such as a mirror or chrome-plated wrench, over the collector
exit. Steady condensation indicates that the flame is lit.
Advanced User Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents