Setting Parameters For The Fpd - Agilent Technologies 7820A Advanced User's Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

5
Detectors

Setting parameters for the FPD

WA RN IN G
106
Manual ignition
Press [Front Det] or [Back Det].
1
Scroll to Flame. Press [On/Yes]. The flame ignition sequence
2
begins.
Automatic ignition
If the FPD output with the flame on falls below the flame-off
output plus the Lit offset value, this is interpreted as a flame-out
condition. The FPD runs the flame ignition sequence to relight
the flame. If this fails, it runs the sequence again. If the second
attempt also fails, the detector shuts down all functions except
temperature and makeup gas flow.
Verify that a column is installed or the FPD fitting is plugged
before turning on the air or hydrogen. An explosion may occur if
air and hydrogen are allowed to leak into the oven.
Press [Front det] or [Back det].
1
Set the detector temperature. It must be greater than 120 °C
2
for the flame to light.
Change the hydrogen flow rate, if desired. Press [Off/No].
3
Change the air flow rate, if desired. Press [Off/No].
4
5
If you are using packed columns, turn off the makeup gas
and proceed to step 9.
If you are using capillary columns:
6
Verify that makeup gas type is the same as that plumbed
a
to your instrument (next to Makeup in the parameter list).
Change the gas type, if necessary.
If your capillary column is defined, choose a flow mode
b
and set the makeup gas flow or combined flow.
c
If your capillary column is not defined, enter a makeup
gas flow. Only constant flow is available.
Scroll to Flame and press [On/Yes]. This turns on the air and
7
hydrogen and initiates the ignition sequence.
Advanced User Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents