Detectors

Selecting an NPD jet

Table 18
Jets for capillary optimized fittings
Figure 3 ID
Jet type
1
Capillary with extended jet
(recommended)
2
Capillary
3
High-temperature
90
Open the oven door and locate the column connection fitting at
the base of the detector. It will look like either a capillary
optimized fitting or an adaptable fitting.
Capillary optimized fitting
• If you have an application that tends to clog the jet, select a
jet with a wider tip id.
• When using packed columns in high column-bleed
applications, the jet tends to clog with silicon dioxide.
For capillary optimized fittings, select one of the following from
Table
18.
Part number
G1534-80580
G1531-80560
G1531-80620
51.5 mm
1
43 mm
2
43 mm
3
Figure 2
Capillary optimized NPD jets
For the adjustable NPD, select one of the following from
Table
19.
Adaptable fitting
Detector fitting
Adapter
Jet tip id
Length
0.29 mm (0.011 inch)
51.5 mm
0.29 mm (0.011 inch)
43 mm
0.47 mm (0.018 inch)
43 mm
1
2
3
Advanced User Guide