Clock Time Programming; Using Clock Time Events; Programming Clock Time Events; Adding Events To The Clock Table - Agilent Technologies 7820A Advanced User's Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

Clock Time Programming

Using clock time events

Programming clock time events

Adding events to the clock table

Advanced User Guide
Clock time programming allows certain setpoints to change
automatically at a specified time during a 24-hour day. Thus, an
event programmed to occur at 14:35 hours will occur at 2:35 in
the afternoon. A running analysis or sequence has precedence
over any clock table events occurring during this time. Such
events are not executed.
Possible clock time events include:
• Valve control
• Method and sequence loading
• Starting sequences
• Initiating blank and prep runs
• Column compensation changes
• Adjustments of the detector offset
The Clock Table function allows you to program events to occur
during a day based on the 24-hour clock. Clock table events that
would occur during a run or sequence are ignored.
For example, the clock table could be used to make a blank run
before you even get to work in the morning.
Press [Clock Table].
1
Press [Mode/Type] to see the available clock time events.
2
Scroll to the parameter to be programmed.
3
Edit Time: and the setpoints for this event.
4
Press [Mode/Type] to add another event. Press [Status] to
5
terminate entries.
When the clock event is executed, a confirming message
appears.
Press [Clock Table].
1
Press [Mode/Type]. When entries are added, they are
2
automatically ordered chronologically.
2
Programming
15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents