3
Inlets
Carrier Gas Flow Rates
22
The flow rates in
Table 2
temperatures.
Table 2
Column size and carrier flow rate
Column type
Column size
Packed
1/8-inch
1/4-inch
Capillary
0.05 mm id
0.10 mm id
0.20 mm id
0.25 mm id
0.32 mm id
0.53 mm id
are recommended for all column
Carrier flow rate, mL/min
Hydrogen
Helium
30
60
0.5
0.4
1.0
0.8
2.0
1.6
2.5
2.0
3.2
2.6
5.3
4.2
Nitrogen
20
40
n/a
n/a
0.25
0.5
0.75
1.5