Setting Parameters For The Npd - Agilent Technologies 7820A Advanced User's Manual

Gas chromatograph
Hide thumbs Also See for 7820A:
Table of Contents

Advertisement

5
Detectors

Setting parameters for the NPD

WA RN IN G
88
gases, including the detector hydrogen, air, and makeup gases.
Do not use plastic (including PTFE) tubing, plastic-bodied traps,
or O-ring seals.
Before operating the NPD, make sure that detector gases are
connected, a column is installed, and the system is free of leaks.
Set the oven temperature, inlet temperature, and column flow.
Make sure that a column is installed or the NPD column fitting is
plugged before turning on the air or hydrogen. An explosion may
occur if air and hydrogen are allowed to leak into the oven.
Select the bead (white ceramic, black ceramic).
1
Select the jet.
2
Install bead and jet as required. (See the Maintenance
3
Manual for details.)
Press [Config][Front Det] or [Config][Back Det].
4
5
If you are using makeup gas, verify that the configured
makeup gas type is the same as that plumbed to your
instrument. Change the gas type, if necessary. Nitrogen is
recommended.
If the displayed Bead Type is incorrect, set the Bead Type
6
using the [Mode/Type] key.
Set Auto Adjust (On recommended).
7
Set Dry Bead (On recommended).
8
Press [Front Det] or [Back Det].
9
Set the detector temperature. The recommended range is
10
325 to 335 °C.
Enter a hydrogen flow (3.0 mL/min is recommended). Turn
11
the flow On.
Enter an air flow (60 is recommended for ceramic beads).
12
Turn the flow On.
a
If you are using packed columns, turn off makeup gas and
proceed to step 13.
If your capillary column is defined, choose a flow mode
b
and set the makeup gas flow. For a column in the constant
Advanced User Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents