Before You Start

Before You Start
There are three things you should do before starting an adjustment
procedure.
• Check that you are familiar with the safety symbols marked on the
analyzer, and read the general safety considerations and the symbol
definitions given in the front of this service guide.
• Check that the analyzer has been turned on and allowed to warm up
for at least 30 minutes at room temperature before making any
adjustments. The analyzer must be allowed to stand at room
temperature at least 2 hours prior to the 30 minute warmup.
• Read the rest of this section.