Agilent Technologies 8590L Service Manual page 605

8590 e-series and 8590 l-series
Table of Contents

Advertisement

Softkey Descriptions
Softkey Descriptions
Enter the new correction constants into A16 RAM with
the following procedure.
1. Press the following keys.
2. The data-entry instruction for the 1 dB step, ENTER
CAL ATTEN ERROR 1, is displayed in the
active-function block, just below the active entry for
the reference-level-offset value, REF LVL
OFFSET.
3. Enter the characterization data for the 1 dB, 2 dB,
4 dB, and 8 dB steps, with a 0.01 dB resolution for
each entry.
4. Terminate each entry with either the
key.
5. Enter the original factory value, or a default value of
0 dBm, for the 16 dB step.
After each entry the characterized value is
displayed, with 0.1 dB resolution, to the left of the
graticule, followed by the offset-reference-level
indicator, OFFSET.
If desired, use
attenuator-characterization data with 0.01 dB
resolution. Refer to
explanation of the displayed calibration data.
R
S
ELATED
OFTKEYS
DISPLAY CAL DATA
606
PRESET
, −2001,
FREQUENCY
SET ATTN ERROR
DISPLAY CAL DATA
DISPLAY CAL DATA
CAL AMPTD
Hz
or −dBm
+dBm
to review the new
for an