Making Adjustments
5. Crystal and LC Bandwidth Filter
15.Remove the crystal shorts and press the following analyzer keys.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF
, 50,
SPAN
kHz
MKR FCTN, MK TRACK ON OFF
, 3,
BW
kHz
PEAK SEARCH, MARKER DELTA
, 30,
BW
kHz
PEAK SEARCH
16.Verify that the MARKER ∆ frequency does not exceed 3 kHz.
If the signal shift is out of tolerance, repeat
17.Press the following analyzer keys.
MKR
MARKER 1 ON OFF
52
(ON)
(OFF)
(OFF)
step 3
to
step
16.