Agilent Technologies 8590L Service Manual page 246

8590 e-series and 8590 l-series
Table of Contents

Advertisement

Verify the −10 V reference detector
This service-diagnostic routine verifies the −10 V reference. −10 V is
summed with the tuning voltages and used in the main coil driver and
originates on the A7 analog interface board assembly. Frequency and
amplitude errors will occur if this voltage is incorrect.
−10 V Reference Detector
Figure 4-14
Chapter 4
Verify the −10 V reference detector by pressing
• If the signal trace is at the bottom of the display, the −10 V
reference detector is functioning properly. See
• If the signal is not at the bottom of the display, the problem is
isolated to the A7 analog interface board assembly. Refer to
Chapter 6
for further troubleshooting.
Troubleshooting the Analyzer
Using the Internal Service-Diagnostic Routines
−10V REF DETECTOR
Figure
.
4-14.
247