Analyzer Messages
Interpreting Analyzer Messages
INVALID ITEM: _ _ _(U)
Indicates an invalid parameter has been used in a programming
command.
INVALID KEYLBL: _ _ _(U)
Indicates that the specified key label contains too many characters. A
key label is limited to 8 printable characters per label line.
INVALID KEYNAME:_ _ _ (U)
The specified key name is not allowed.
Related Assemblies
A16 processor/video assembly.
Troubleshooting Hints
The key name may have conflicted with an analyzer programming
command. To avoid this problem, use an underscore as the second
character in the key name, or avoid beginning the key name with the
following pairs of letters: LB, OA, OL, TA, TB, TR, MA, MF, TS, OT, and
DR.
INVALID OUTPUT FORMAT (U)
The output format is not valid. See the appropriate programming
command description to determine the correct format.
INVALID RANGE: Stop < Start (U)
Indicates that the first trace element specified for a range of trace
elements is larger that ending trace element. When specifying a trace
range, the starting element must be less than the ending element. For
example, TRA[2,300] is legal but TRA[300,2] is not.
INVALID REGISTER NUMBER (U)
The specified trace register number is invalid.
INVALID REPEAT MEM OVFL (U)
Memory overflow occurred due to a REPEAT routine. This can occur if
there is not enough analyzer memory for the REPEAT UNTIL
declaration, or if the REPEAT UNTIL declaration exceeds 2047
characters.