Procedure
It is recommended to complete the "CAL FREQ Adjustment Routine"
prior to performing the "CAL AMPTD Adjustment Routine."
1. Connect the CAL OUT to the INPUT 50 Ω using a BNC cable. See
Figure 2-14
CAL AMPTD Adjustment Routine Setup
Use only 75 Ω cables, connectors, or adapters on instruments equipped
CAUTION
with 75 Ω inputs or damage to the input connectors will occur.
2. Press the following analyzer keys.
Chapter 2
Figure
2-14.
75 Ω input: Connect the CAL OUT to the INPUT 75 Ω with the 75 Ω
BNC cable.
CAL, CAL AMPTD
The
routine takes approximately 5 to 7 minutes to run.
CAL AMPTD
The internal adjustment data will be stored in working RAM. To
store this data in nonvolatile memory, press
Making Adjustments
9. CAL AMPTD Adjustment Routine
.
CAL STORE
71