Agilent Technologies 8590L Service Manual page 646

8590 e-series and 8590 l-series
Table of Contents

Advertisement

MEAS UNCAL (U)
The measurement is uncalibrated. Check the sweep time, span, and
bandwidth settings, or press
No card found (U)
Indicates that the memory card is not inserted.
No points defined (U)
Indicates the specified limit line or amplitude correction function
cannot be performed because no limit line segments or amplitude
correction factors have been defined.
MIXER BIAS CAL FAILED
This message applies to 8592L, 8593E, 8595E, and 8596E spectrum
analyzers only.
During the
CAL MXR
displayed center frequency.
Related Assemblies
A3A1 comb generator, A3A11 step-recovery diode, A3A12 3dB
attenuator, J2 connector.
Troubleshooting Hints
• Make sure that the 100 MHz COMB OUT is connected to the
analyzer input using the comb generator cable (SMA to Type N)
supplied with the analyzer. (A comb generator cable (SMA to SMA)
is supplied with Option 026 instruments.)
• Perform the following steps to verify the comb-generator output
manually.
1. Press the following keys.
PRESET
FREQUENCY
500,
SPAN,
MHz
2. Look for the 100 MHz comb-generator signal to the right of the
local oscillator signal. When
signal has an amplitude of approximately −8 dBm and the
100 MHz harmonics are less than −20 dBc.
3. Press the following keys.
AUX CTRL
COMB GEN ON OFF
Chapter 14
AUTO COUPLE, AUTO ALL.
routine, the comb signal was not found at the
, 100,
MHz
COMB GEN ON OFF
(On)
Analyzer Messages
Interpreting Analyzer Messages
is OFF, this
647

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

8591e8591c8594e8594q8594l8595e ... Show all

Table of Contents