Special Calibrations

Perform a "Special" Calibration
CHSE?
Triggers special calibration
ADC
Aligns the different resolutions available from the analog to digital converter, so that
there are no significant differences in readings when changing resolutions with a constant
input value.
FREQ
Calibrates the frequency counter against an external frequency standard, by correcting an
internal frequency 'gain' factor.
HV_LIN
ONLY used after the five linearity measurements have been taken using the command
HV_LIN? <parameter>, to calculate the DC high voltage linearity correction from these
measurements.
Response
(Nr1) The value returned identifies the success (0) or failure (1) of the calibration step.
Execution Errors
If special calibration is not enabled.
"Special" measurement commands for DCV Linearity Calculations
HV_LIN?
"Special" measurement commands for DCV Linearity Calculations
HV_LIN?
Triggers a measurement and records the result for use with the special calibration
command CHSE? HV_LIN. The signal supplied for each measurement should match the
supplied parameter.
SRC_N1K
Measurement source is -1000V DC.
phs
CHSE?
Special Calibration
phs
Remote Operations Using the IEEE 488 Interface
Calibration Commands and Messages
A DC
FREQ
HV_LIN
S R C _ N 1 K
SRC_N500
ZERO
SRC_P500
SRC_P1K
4
adj280f.eps
adj361f.eps
4-49