8508A
Users Manual
8508A Range
2 Ω
20 Ω
200 Ω
2 kΩ
20 kΩ
200 kΩ
2 MΩ
20 MΩ
200 MΩ
2 GΩ
20 GΩ
1. 20 GΩ range gain is adjusted automatically during 2 GΩ HiVΩ range gain adjustment, but may be adjusted
separately (after 2 GΩ HiVΩ range adjustment) at 10 GΩ if a standard is available.
6-20
Table 6-5. Resistance Adjustment Points and Sequence
Modes
Normal, Normal LoI,
TruΩ, TruΩ LoI
Normal, Normal LoI,
TruΩ, TruΩ LoI
Normal, Normal LoI,
TruΩ, TruΩ LoI
Normal, Normal LoI,
TruΩ, TruΩ LoI
Normal, Normal LoI,
TruΩ, TruΩ LoI
Normal, Normal LoI
Normal, Normal LoI
Normal, Normal LoI,
HiVΩ
Normal, Normal LoI,
HiVΩ
Normal, Normal LoI,
HiVΩ
HiVΩ
Point
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Zero
Range Gain
Resistance
0 Ω
1 Ω
0 Ω
10 Ω
0 Ω
100 Ω
0 Ω
1 kΩ
0 Ω
10 kΩ
0 Ω
100 kΩ
0 Ω
1 MΩ
0 Ω
10 MΩ
0 Ω
100 MΩ
0 Ω
1 GΩ
0 Ω
1
Optional