Fluke 8508A User Manual page 123

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

Fourth field -
Options
*OPT?
(String) Recalls the instrument's option configuration which matches the options
provided in the sales literature.
Example:
"8508A/01" -- for instrument fitted with rear terminals.
Setting Line Frequency
Selection of line frequency setting adjusts the integration time of the A-D converter for
improved line frequency noise rejection.
LINEF 50
Selects a line frequency operation of 50Hz.
LINEF 60
Selects a line frequency operation of 60Hz.
The selection is saved in non volatile memory. Note The only allowed values of
for 50Hz, and
50
will be rounded to that resolution.
Power On and Reset Conditions
Reactivated at Power On
Recall of Line Frequency Setting
LINEF?
(Nr1) Recalls the active setting for line frequency.
Firmware level (issue.revision).
*
OPT?
Options
LINEF
phs
LINEF
phs
Setting Line Frequency 50 Hz or 60 Hz
for 60Hz. Numbers exceeding the defined data element resolution
60
LINEF?
Recall Line Frequency Setting
Remote Operations Using the IEEE 488 Interface
50
60
Instrument I/D and Setup
adj270f.eps
adj281f.eps
are
Nrf
adj282f.eps
4
4-45