Interconnections - Fluke 8508A User Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Users Manual
Enable Calibration
Enter Cal mode
HVlin Cal
Adc Cal
DCV
ACV
DCI
ACI
Frequency
Ohms
Exit Cal mode
NV input offsets
Set CAL DUE
Disable Calibration
1.
To meet user's need, just one range on one function can be adjusted.

Interconnections

The Fluke 8508A-7000K Calibration kit contains a specially designed lead set which
connects directly to the terminals of the 8508A, the 5720A, and the 5725A for the
purpose of 8508A calibration. It is recommended that this lead kit be used to avoid
introducing measurement errors due to interconnections and to duplicate the
configuration used by Fluke for 8508A calibration. Figure 6-1 shows the interconnections
used in detail. If alternate leads are substituted, make sure they are metrology grade leads,
and of adequate voltage and current ratings. Leads should be kept as short as possible,
constructed from low capacitance low leakage dielectric material (PTFE) with low
thermal emf terminations, and exactly duplicate the configuration and interconnections
shown.
When adjusting the 2 GΩ and 20 GΩ resistance ranges the 1 GΩ standard resistor
supplied in the Fluke 8508A-7000K Calibration kit plugs directly into the 8508A front
input terminals, without any interconnecting leads. If alternate resistance standards are
used for resistance adjustment, refer to the resistance-measurements section of Chapter 3
Making Measurements for connection details.
The Fluke 8508A-7000K Calibration kit also contains two precision low-thermal emf
4-wire shorting devices for use when performing the non-volatile input offset adjustments
at the front and rear input terminals. These devices link the input terminals in the
following order: Input Lo to Input Hi to Sense Hi to Sense Lo. If these devices are not
available it is recommended that a U-shaped piece of thick bare copper wire be used to
link the terminals in the same manner.
6-12
Table 6-2. Adjustment Sequence
General Sequence for Full Instrument Adjustment
Set rear panel calibration switch to ENABLE.
Access CALIBRATION menu .
Corrects linearity of the 1 kV DC range. Performed only if required.
Analog to digital converter alignment. Performed only if required.
Range Zeros and Range Gains (200 mV to 1 kV Ranges).
Range Zeros, Range LF Gains and Range HF Gains (200 mV to 1 kV
Ranges).
Range Zeros and Range Gains (200µA to 20A Ranges).
Range Zeros and Range Gains (200µA to 20A Ranges).
Frequency counter cal at 1MHz
Range Zeros and Range Gains:
Ohms: 2 Ω to 2 GΩ Ranges, then LoI Ohms.
TruΩ: 2 Ω to 20 kΩ Ranges, then LoI TruΩ.
HiVΩ: 20 MΩ to 20 GΩ Ranges.
Quit the CALIBRATION menu.
Front & rear non-volatile input offset adjustments:
DCV, Ohms, TruΩ, HiVΩ (including LoI , 2 & 4-wire).
Set the date for the next adjustment