Reference Multimeter and 8 ½ Digit Multimeter
Triggering Measurements

Special Event Qualifiers

Two special qualifiers can be sent remotely to bypass the Trigger Event Detector
irrespective of the Event specified by the 8 trigger events described above. (The
8 trigger events are parameters of the TRIGger:SOURce command.) These
qualifiers are not available from the front panel (in Local control). The two ARM
layers also implement the same special qualifiers. See Figure 29.
TRIG:SIGNal
TRIG:IMM
iei339.emf
Figure 29. Trigger Layer with Special Events
TRIGger:SIGNal causes immediate exit from the trigger layer Event Detector.
The process flow rejoins immediately below the Event Detector. TRIGger:SIGNal
is a one-time command that does not modify any settings. The ARM2 command
is ARM2:SIGNal, and ARM1 is ARM1:SIGNal.
TRIGger:IMMediate causes immediate exit from the trigger layer Event Detector.
The process flow rejoins immediately below the Delay block. TRIGger:IMMediate
is a one-time command that does not modify any settings. The ARM2 command
is ARM2:IMMediate, and ARM1 is ARM1:IMMediate.