Table 6-9. AC Voltage Verification Points and Sequence
8508A Range
Voltage
200 mV
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
2 V
20V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
19 V
Frequency
20 Hz
55 Hz
1 kHz
3 kHz
10 kHz
30 kHz
60 kHz
100 kHz
1 V
20 Hz
1 V
55 Hz
1 V
1 kHz
1 V
3 kHz
1 V
10 kHz
1 V
30 kHz
1 V
60 kHz
1 V
100 kHz
1 V
500 kHz
1 V
1 MHz
1 V
1 kHz
20 Hz
55 Hz
1 kHz
3 kHz
10kHz
30 kHz
60 kHz
100 kHz
500 kHz
1 MHz
1 kHz
Adjustment and Verification
Performance Verification
8508A Filter
Tolerance
± 0.0195 mV
10 Hz
± 0.0175 mV
40 Hz
± 0.0149 mV
40 Hz
40 Hz
± 0.0175 mV
40 Hz
± 0.0175 mV
40 Hz
± 0.0445 mV
± 0.0995 mV
40 Hz
± 0.0995 mV
40 Hz
± 0.000144 V
10 Hz
± 0.000119 V
40 Hz
± 0.000099 V
40 Hz
± 0.000119 V
40 Hz
± 0.000119 V
40 Hz
40 Hz
± 0.000295 V
40 Hz
± 0.000795 V
40 Hz
± 0.000795 V
± 0.034000 V
40 Hz
± 0.034000 V
40 Hz
± 0.00032 V
40 Hz
± 0.00144 V
10 Hz
± 0.00119 V
40 Hz
± 0.00099 V
40 Hz
± 0.00119 V
40 Hz
± 0.00119 V
40 Hz
40 Hz
± 0.00295 V
40 Hz
± 0.00795 V
40 Hz
± 0.00795 V
± 0.34000 V
40 Hz
± 0.34000 V
40 Hz
± 0.00167 V
40 Hz