Repair and maintenance
11.1 Basic safety instructions
Maintenance during continued operation in a hazardous area
There is a danger of explosion when carrying out repairs and maintenance on the device in
a hazardous area.
• Isolate the device from power.
- or -
• Ensure that the atmosphere is explosion-free (hot work permit).
Commissioning and operation with pending error
If an error message appears, correct operation in the process is no longer guaranteed.
• Check the gravity of the error.
• Correct the error.
• If the error still exists:
See also
Display in case of a fault (Page 185)
Hot, toxic or corrosive process media
Danger of injury during maintenance work.
When working on the process connection, hot, toxic or corrosive process media could be
released.
• As long as the device is under pressure, do not loosen process connections and do not
• Before opening or removing the device ensure that process media cannot be released.
Improper connection after maintenance
Danger of explosion in areas subject to explosion hazard.
• Connect the device correctly after maintenance.
• Close the device after maintenance work.
Refer to Chapter "Connecting up (Page 63)".
182
WARNING
WARNING
– Take the device out of operation.
– Prevent renewed commissioning.
WARNING
remove any parts that are pressurized.
WARNING
SITRANS P DS III/P410 with PROFIBUS PA
Operating Instructions, 02/2016, A5E00053276-08