Performance Characteristics Vertical System - HP 54616B User And Service Manual

Table of Contents

Advertisement

Sensitivity 2 mV/div to 5 V/div
1
±2.0% of full scale
Accuracy
1
Fully calibrated, accuracy ±2.0 % of full scale
Verniers
1, 2, 3
Cursor accuracy
Single cursor accuracy: vertical accuracy ±1.2% of full scale ±0.5% of
position value
Dual cursor accuracy: vertical accuracy ±0.4% of full scale
Bandwidth limit ≈30 MHz
Coupling Ground, ac, and dc
Inversion Channel 1 and channel 2
CMRR (common mode rejection ratio) ≥ 20 dB at 50 MHz
Probe Sense Automatic readout of 1X, 10X, 20X and 100X probes
50Ω protection Protects 50Ω load from excessive voltage.
Time skew Adjustable over a range of ±25 ns to remove effects of cabling
and probe delays.
1
When the temperature is within ±10 °C from the calibration temperature.
2
Magnification is used below 7 mV/div range. Below 7 mV/div full scale is
defined as 56 mV.
3
Tested, see "To verify voltage measurement accuracy" on page 4-8.
Performance Characteristics
Vertical System
5-3