Acquisition System

Maximum sample rate
HP 54615B – 1 GSa/s simultaneous on 2 channels
HP 54616B/16C – 2 GSa/s simultaneous on 2 channels
Resolution 8 bits
Simultaneous channels Channels 1 and 2
Record length
Vectors off:
5000 points
4000 points (200 ns/div, HP 54615B)
4000 points (100 ns/div, HP 54616B/16C)
1000 points (peak detect on)
Vectors on:
2000 points
1000 points (peak detect on)
Roll Mode (vectors off or on):
1000 points
500 points (200 ms/div, channel 1 and 2 on, HP 54615B/16B)
Single-shot bandwidth
HP 54615B – 250 MHz on channels 1 and 2 simultaneously (1 GSa/s,
display vectors on)
HP 54616B/16C – 500 MHz on channels 1 and 2 simultaneously
(2 GSa/s, display vectors on)
Acquisition modes Normal, Peak Detect, and Average
Peak detect 1 ns glitch capture
Average Number of averages selectable at 8, 64, and 256
Roll Mode At sweep speeds of 200 ms/div and slower (HP 54615B/16B)
At sweep speeds of 500 ms/div and slower (HP 54616C):
waveform data moves across the display from right to left with no
dead time. Display can be either free-running (non-triggered) or
triggered to stop on a trigger event.
Performance Characteristics
Acquisition System