Performance Characteristics

Horizontal System

Horizontal System
Sweep speeds 5 s/div to 1 ns/div main and delayed
Accuracy ±0.005% of reading
Horizontal resolution 20 ps
1
(∆t and 1/∆t) ±0.005% ±0.2% of full scale ±100 ps
Cursor accuracy
Delay jitter ≤1 ppm
Pretrigger delay (negative time)
HP 54615B–The greater of 30 µs or 60 divisions, not to exceed 100 s
HP 54616B/16C–The greater of 15 µs or 60 divisions, not to exceed 100 s.
Posttrigger delay (from trigger point to start of sweep)
The greater of 10 ms or 20,000 divisions, not to exceed 100 s.
Delayed sweep operation
Up to 200 times main sweep when main sweep is from 5 s/div to 10 ms/div.
Up to 1 ns/div with main sweep set to 5 ms/div and faster.
Horizontal modes Main, Delayed (Alt), X-Y, and Roll
1
Tested, see "To verify horizontal ∆t and 1/∆t accuracy," on page 4-13.