HP 54616B User And Service Manual page 95

Table of Contents

Advertisement

Table 4-1
Recommended list of test equipment to service the oscilloscope
Equipment
Critical specifications
Signal generator
1 to 500 MHz at 200 mV
high stability timebase
Digital multimeter
0.1 mV resolution, better than 0.01%
accuracy
Oscilloscope
100 MHz, 1 MΩ input R
1 to 500 MHz ±3% accuracy
Power meter and
Power sensor
Power supply
14 mV to 35 Vdc, 0.1 mV resolution HP 6114A
Rise time ≤ 700 ps
Pulse generator
Power splitter
Outputs differ < 0.15 dB
Shorting cap
BNC
Time Mark
Stability 5 ppm after 30 minutes
Generator
2
Dummy load
Compatible with power supply
Adapter
SMA (f) to BNC (m)
Adapter
BNC (f-f)
Adapter
BNC tee (m) (f) (f)
Adapter
N (m) to BNC (f), Qty 3
Adapter
BNC (f) to dual banana (m)
Adapter
Type N (m) to BNC (m)
Cable
BNC, Qty 3
Cable
BNC, 9 inches, Qty 2
Cable
Type N (m) 24 inch
3
Adapter
BNC (m) to dual banana post
1
P = Use for Performance Verification.
A = Use for Adjustments.
T = Use for Troubleshooting.
2
See page 4-33 to construct your own dummy load.
3
Used for Option 005 only
Recommended Model/Part
HP 8656B Option 001
HP 34401A
HP 54600
HP 436A and HP 8482A
HP 8131A
HP 11667B
HP 1250-0774
Tektronix TG501A and TM503B
see note 2 below
HP 1250-1787
HP 1250-0080
HP 1250-0781
HP 1250-0780
HP 1251-2277
HP 1251-0082
HP 10503A
HP 10502A
HP 11500B
HP 10110B
Service
1
Use
P
P, A, T
P, T
P
P
A
P
P
P
A
A
P, A
P, A
P
P
P
P, A
P, A
P
P
4-3