HP 54616B User And Service Manual page 120

Table of Contents

Advertisement

Service
Adjusting the Oscilloscope
Adjust the channel 1 high-frequency pulse response for an %
10
overshhot of 5% mininum to 7% maximum.
Repeat steps 2 through 10 substituting Channel 2 for Channel 1 and
11
∆V(2) for ∆V(1).
Figure 4-3
High-frequency pulse
response adjustments
High-frequency pulse response adjustment locations
4–28
If a low loss 50Ω cable is used (such as HP 8120-4949), the % overshoot should
be 6% minimum to 8% maximum.)