2.4 Special commissioning function

2.4.1 Device software and checksum

Move the cursor under tab "Serv." to
The checksum

2.4.2 Clear volumes and archives ("Clr.V")

i
The
"Clr.V"
So that all volume counters and all archives and logbooks are deleted from the device, open the
calibration lock as described in the EK280 operating instructions.
Move the cursor in tab
Press the ENTER button 
Change the value using the arrow buttons  or  to "1".
Press the ENTER button to confirm the set value. It is possible to abort the entry by pressing the
ESC button.

2.4.3 Clear archives ("Clr.A")

i
With the
"Clr.A"
monthly archives 1 and 2 and the four user archives can be deleted.
So that the measuring period archive, the daily archive, monthly archives 1 and 2 and the four user
archives are deleted from the device, go to and open the administrator or calibration lock as
described in the EK280 operating instructions.
Move the cursor in tab
Press the ENTER button  One digit of the value flashes.
A safety function is provided to prevent accidental deleting of meter counts and archives. The serial
number of the unit must be entered to activate the safety function. For input, move the cursor with
the arrow keys  or  to the digits and change these with the arrow keys  or .
i
The serial number of the device is printed on the identification label or it can be fined in the
register under
Press the ENTER button to confirm the value setting. Input can be canceled using the ESC key.

2.4.4 Parameterize the volume at measurement conditions "Vm"

i
A one-off setting of the volume counter is possible from software version 2.20 while the
administrator lock is open.
The counter of the EK280 can be set once in a similar way to that of the gas counter while the administrator
lock is open for recording the volume in practice. When the calibration lock is open it is always possible to
set the volume and this can also be done directly via the operational volume (tab
"Actual volume" >
"Vm".
"Chk"
can be recalculated for verification purposes by pressing the ENTER button.
functions described here delete all volume counters and all archives and logbooks.
"Edit and delete" > "Clr.V".
"Serv."
to
"0"
flashes.
functions described here, only the measurement period archive, the daily archive,
"Serv."
to
"Edit and delete > "Clr.A"
"Identification" > "Volume Converter" >
Application Manual
"Identification" > "Volume Converter" > "Vers"
"SNo".
.
"Serv."
11
or
"Chk"
"Serv."
to
"Volume" >