Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals
Guidance and manufacturer's declaration IEC 60601-1-2: Edition 4.0; 2014-02 – Electromagnetic immunity
Icare HOME (TA022) is intended for use in a home healthcare environment with
electromagnetic characteristics
the Icare HOME (TA022) should assure that it is used in such an environment.
IEC 60601 test
Immunity test
level
Conducted
3 V
disturbances
0,15 MHz – 80
induced by RF
MHz
fields
IEC 61000-4-6
6 V in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz – 2,7 GHz
80% AM at 1 kHz
specified below. The customer or the user of
Compliance
Electromagnetic environment-Guidance
level
3 V
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
6 V
part of the Icare HOME (TA022), including
cables specified by the manufacturer.
Otherwise, degradation of the performance of
this equipment could result.
10 V/m
WARNING: Portable RF communications
equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to any
part of the Icare HOME (TA022) including cables
specified by the manufacturer. Otherwise,
degradation of the performance of this
equipment could result.
Interference may occur in the vicinity of
equipment marked with the following symbol:
ENGLISH