Icare HOME tonometer Instruction manual for health care professionals
Icare HOME tonometer Instruction manual for health care professionals
Guidance and manufacturer's declaration IEC 60601-1-2: Edition 4.0; 2014-02 - Electromagnetic immunity
Icare HOME (TA022) is intended for use in a home healthcare environment with electromagnetic
characteristics specified below. The customers or users of Icare HOME (TA022) should assurre that it is
used in such enviroment
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
Transients / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruption and voltage
variations on power
supply lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field IEC
61000-4-8
IEC 60601 Test level
± 8 kV contact
± 2kV, ±4kV, ±8kV, ±15 kV air
± 2 kV
100 kHz repetition
frequency
±1 kV for line(s) to line(s)
±2 kV for line(s) to earth
0 % UT for 0.5 cycle
(1 phase)
0 % UT for 1 cycle 70
% UT for 25/30
cycles (50/60 Hz) 0 %
UT for 250/300
cycles
(50/60 Hz)
30 A/m
ENGLISH
Electromagnetic
Compliance level
environment-Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
± 8 kV contact
the relative humidity should be at
± 15 kV air
least 30%
Icare HOME (TA022) tonometer
is not operational when
connected to an external
computer, Icare HOME (TA022)
Not applicable
batteries are not rechargeable
Icare HOME (TA022) tonometer
is not operational when
connected to an external
Not applicable
computer, Icare HOME (TA022)
batteries are not rechargeable
Icare HOME (TA022) tonometer
is not operational when
connected to an external
Not applicable
computer, Icare HOME (TA022)
batteries are not rechargeable
30 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
WARNING: Sources of power
frequency magnetic field should
be used no closer than 15 cm (6
inches) to any part of Icare
HOME (TA022), including cables
specified by the manufacturer.
Otherwise, degradation of the
performance could result.
www.icare-usa.com
31