STEP 5 – Test measurement
1. Load a new probe and ask the same HCP to measure his/her own IOP three times in
each eye with the same tonometer used in the reference measurement.
2. Observe if the positioning is correct. Do not supervise or interact.
STEP 6 – Certification
Connect device to computer and view the readings in the Icare LINK software.
The HCP passes the training and is certified for performing tonometry on
patients with the device and for training others in self-use of the device if the
following conditions are met:
a. The reading taken by the trainer and the first of the three readings taken by
the HCP differ by 5 mmHg or less.
b. The range (max-min) of the three readings taken by the HCP is 5 mmHg or
less in case the first reading is 7-23 mmHg or 7 mmHg or less in case the first
reading is >23 mmHg.
c. The positioning of the tonometer was correct during self-use as determined
by the trainer.
HCP's name:
CERTIFICATION
Icare HOME
0 = Pass; 1 = Fail
ENGLISH
Date (dd mm yy):
37