9.4

Calibration of analog out (DTI 1000 only)

Use following procedure for recalibration of the two analog outputs:
1.
Select the menu analog out.
2.
Enter the date format as YYMMDD.
3.
Connect the voltmeter to channel 1 as shown in the figure section 5.2.2. Read the
voltage and enter the measured value for channel 1.
4.
Connect the voltmeter to channel 2 as shown in the figure section 5.2.2. Read the
voltage and enter the measured value for channel 2.
Enter - Analog output channel 1 :
Enter - Analog output channel 2 :
< ESC > aborts
5.
Connect the voltmeter to channel 1 again, and enter the measured signal for
channel 1.
6.
Connect the voltmeter to channel 2 again, and enter the measured signal for
channel 2.
Enter - Analog output channel 1 :
Enter - Analog output channel 2 :
< ESC > aborts
7.
Enter the analog output scaling (Default: 5mV/°C, 0V=0°C).
100975
05
Lowest possible voltages
Highest possible voltages
Analog fixpoints
Gain
:
Offset
:
< ESC > aborts
11-06-2001
-0.6
-0.6
3.6
3.6
5
mV/°C
°C
0
Volt
Volt
Volt
Volt
31/32