The edit logdata menu is for removal of unwanted logs where the defined softkeys are
shown in it's own menu-frame:
1
2
3
When the calibration points are measured use <F4> to calculate the new coefficients (C will
only be shown for logs below 0°C). Leave the menu with <ESC> and enter the calculated
coefficients in the menu: "Enter new coefficients".
8.7
Recalibrate (Calibration information)
This menu will show the calibration date if the program is installed on the hard disk.
It is only possible to re-calibrate the DTI when the program is executed from the floppy disk.
Please see section 9 for calibration.
100975
05
Logdata : TEST1.CLO
Sensor 1
Ohm
Read
0.020
0.020°C
100.000
100.000°C
199.998
199.998°C
11-06-2001
True
Deviation
0.000°C
-0.020°C
100.000°C
0.000°C
200.018°C
+0.020°C
27/32