5.3.5

Resolution

In order to change the resolution on the read-out, you must press the softkey <MENU/Enter>
3 times. The following message will be shown in the display:
Use the softkey <Reset/SELECT> to scroll between following types of resolution:
RESOLUTION SELECTION MENU
RESOLUTION: 0.01
│ DTI 1000 only
RESOLUTION: 0.001
│ DTI 100 + 1000
RESOLUTION: 0.1
Use the softkey <Menu/ENTER> to select the specific read-out unit.
The sample period is dependent on the selected resolution which can be longer than sample
rate to the display if a resolution above 0.1 is selected. The sample rates for the different
resolutions are shown below:
Resolution
0.1
Sample period
1 s
Please note that read-out in Ω has one more digit than specified in the resolution set-up.
100975
05
RESOLUTION: 0.01
DTI 100 + 1000
0.01
2 s
11-06-2001
READ-OUT MODE
23.65 °C
T1
23.60 °C
T2
T1
23.651°C
T2
23.599°C
T1
23.7
T2
23.6
0.001
10 s
°C
°C