Checking Sensor Coefficients; Light - Ametek DTI 100 User Manual

Digital temperature indicator
Table of Contents

Advertisement

5.3.6

Checking sensor coefficients

In order to check the sensor ID and coefficients on the DTI, press the softkey
<MENU/Enter> 4 times. The following message will be shown in the display:
Use the softkey <Reset/SELECT> to scroll through the coefficients:
(example with both sensors using default coefficients)
SENSOR 1: 9402244
R0
:
1.0000000E+02
SENSOR 1: 9402244
A
:
3.9080200E-03
SENSOR 1: 9402244
B
: -5.8020000E-07
SENSOR 1: 9402244
C
: -4.2735000E-12
Use the softkey <Menu/ENTER> to leave this menu.
The coefficients are used in the equations:
Below 0°C:
Above 0°C:
5.3.7

Light

By pressing the softkey <Light> 1 time, the light will be switched on in the display for about
10 seconds. By pressing the softkey 2 times in a row the light will stay on. The light is
switched off by pressing the softkey again.