Power-On Routine and Error Messages

Once your have turned on the balance using the
the balance is ready to operate.

Error Messages

On the MC 5, error messages are indicated by a code number.
All codes for operating errors, including errors in the sequence of a function, have two-digit numbers.
Error codes for hardware errors have three-digit numbers.
The error codes enable you to identify the type of operating error.
Display
Meaning
Err 01
Display format overflow
Err 02
Zero point error at the start
of the CAL routine
Err 04
Control error during
multiple calibration mode
Err 06
Internal calibration weight
defective or not available
Err 10
1st tare memory disabled
when a value is stored in
2nd tare memory
Err 11
Attempt to store an invalid
number in tare memoryfor example -
using TAR -F1/
Err 20
The ></ key was pressed
before an application
program was initialized
Err 21
Reference percentage or
reference piece count setting
disabled, or this application
does not have a parameter mode
Err 22
Reference cannot be stored
for weighing in percent or counting
Err 23
Value cannot be stored
for any particular application program
Err 30
The print key was pressed
in the BPI mode; no printer on-line
Err 50
Converter error (TC range)
Err 53
Temperature sensor circuit failure
Err 54
Balance converter value below
the minimum limit
36
key, " 0.000 mg " is displayed. This indicates that
Reason/Solution
The weighing system is overloaded.
Re-tare;
check system.
Check system for stability.
Check system. Return the microbalance
to the factory, if necessary.
The tare functions are
interlocked. Clear tare
memory to enable the CF/ tare key.
The readout was negative,
check sample.
For instance, you forgot
to store a reference
percentage for weighing in percent.
Code 2 3 1 is selected in the balance
operating menu. Change this code
to 2 3 2 or 2 3 3.
The weight is too light or
no sample is on the pan.
No sample is on the pan or
the readout is negative.
Connect printer.
Check system or return microbalance to the
factory in Goettingen.
Check temperaturecompensating sensor or
return microbalance to the factory in Goettingen.
Check system.
SARTORIUS MC 5 Balance
Beam at lower stop.
Stand
1/94