Profile Control / Digital Inputs; Figure 14: External Profile Control - Tescom ER3000 User Manual

Electronic pressure controller
Hide thumbs Also See for ER3000:
Table of Contents

Advertisement

4.5.4 Profile Control / Digital Inputs

This feature is only available on models that have the extra analog inputs, as
referenced in section 3 (ER3000 Model Descriptions).
Figure
shows a simple wiring method of controlling a profile, without the
14
need for a computer. Once the profile is downloaded into the ER3000, it can be
started or stopped with a toggle switch. The Digital Input pushbutton may be
used if the "Digital Input" function has been used within the profile. This allows
an operator the flexibility to wait until an event has occurred like changing to the
next device.
An additional feature, necessary for some demanding applications, is provided
for by the Profile Start/Stop and Digital Input signals, which are the Extra
Analog Input 1 and Extra Analog Input 2 inputs of the ER3000. This feature is
the ability to adjust the toggle level (or trip point) of the analog input where it
switches between a logical 0 and a logical 1. The ER3000 toggle variables
enable this feature and are described in section 9.5, Analog Input Variables. For
example, this feature allows connecting a pressure transducer to the Digital Input
and waiting for a specific pressure before continuing on in the pressure profile.

Figure 14: External Profile Control

NOTE: Wire power as shown in figures 2 & 3
(J5 jumpers 3 and 4 must be removed, refer to
)
Figure 1
4. Installation 23