6.6 CLEANING FUNCTION CALIBRATION

1.
[MODE]
D6
2.
[CAL]
Manual cleaning function (MANUAL CLEAN)
The instrument will show the following display:
[MODE]
3A.
[/\] [\/]
4A.
[ENT]
Operator's Manual - Rev. C - 05/05
to select the value
to confirm and to go to the delay time selection
|CAL AL D: 25.0s |
D 25.0 s: actual delay time
to stop the procedure and to go back to D5
to select the value
to confirm and to go back to D5
MESSAGE
| " UPDATE "
to go to
|CLEANING OFF
CLEANING OFF: cleaning function disabled
MANUAL CLEAN: manual cleaning function
AUTO CLEAN: automatic cleaning function
to access to the calibration sequences
(only for MANUAL CLEAN or AUTO CLEAN)
|CLEAN C.:WAITING|
WAITING: the unit is awaiting to start a new Clean Cycle
to stop the procedure and to go back to D6
to select START or WAITING
to confirm selection
- If START is selected the unit go back to D0 and a new Clean Cycle
starts
- If WAITING is selected the unit go back to D6.
_
Delay time calibration
FUNCTION
_
the new data have been memorized
|
_
(MANUAL CLEAN/AUTO CLEAN)
|
_
(START)