6.3 TEMPERATURE CALIBRATION

1.
[MODE]
D2
2.
[CAL]
[MODE]
[/\]+[\/]+[ENT] to press the three keys to turn to factory calibration
3.
[/\] [\/]
4.
[ENT]
[ENT]
Manual Temperature calibration
[MODE]
1.
[/\] [\/]
2.
[ENT]
Operator's Manual - Rev. C - 05/05
to go to
|TEMP.:
20.0 °C |
to access the calibration procedure
|CAL T
20.0°C
'>>>>>' ('<<<<<'): Temperature value overrange
to stop the procedure and to go back to D2
to modify the actual reading
to confirm and to go to the manual Temperature adjustment
MESSAGE
|
" UPDATE "
Error messages
| Z> 2.0°C
The above message will last for 5 minutes
to acknowledge the message
|" NO UPDATE "
|CAL T.M:
20.0°C|
to stop the procedure and to go back to D2
to modify the actual value
to confirm and to go back to D2
_
_
|
FUNCTION
_
The calibration is accepted
|
_
Zero > 2.0 °C
|
_
calibration not accepted
|
_