Set-Point A/B Calibration - B&C CL 7685 Operator's Manual

Residual chlorine chlorine dioxide d. ozone microprocessor controller
Table of Contents

Advertisement

B&C Electronics

6.4 SET-POINT A/B CALIBRATION

The following procedure are suitable for both set-point A and B.
For each set-point it is possible:
-
to insert the set-point value
-
to insert parameters of On/Off - PFM - PWM function
1.
[MODE]
D3
or
D4
2.
[CAL]
Set-point adjustment
[MODE]
3.
[/\] [\/]
4.
[ENT]
A.
On/Off function calibration
B.
PFM function calibration
C.
PWM function calibration
Note: to modify only set-point value, press [ENT] twice until "UPDATE" message
On/Off function
The instrument will show the following display:
Operator's Manual - Rev. C - 05/05
to go to
|SA: 0.60
* F
à
|SB: 0.80
* O
to access the programming sequences
|CAL SA S: 0.60
S 0.60: actual set-point value
to stop the procedure and to go back to D3/D4
to insert the set-point value
- to confirm the value
- to go to one of the following calibration:
|CAL SA I:
20
I 20: actual hysteresis value
_
Set-point A display
L |
_
Set-point B display
H|
ì
_
|
_
Hysteresis calibration
|
CL 7685
-
27