B&C Electronics
PWM function
The instrument will show the following display:
[MODE]
5C.
[/\] [\/]
6C.
[ENT]
[MODE]
7C.
[/\] [\/]
8C.
[ENT]
6.5 ALARM ADJUSTMENT
The following operations are possible:
-
to select the min/max alarm value
-
to select the delay time value
1.
[MODE]
D5
2.
[CAL]
[MODE]
3.
[/\] [\/]
4.
[ENT]
[MODE]
Operator's Manual - Rev. C - 05/05
|CAL SA BP: 0.10 |
BP 0.10: actual proportional band value
to stop the procedure and to go back to D3/D4
to select the proportional band value
to confirm and to go to the selection of the pulse length value
|CAL SA D:
5.0/s|
D 5.0 s: actual pulse length value
to stop the procedure and to go back to D3/D4
to select the pulse length value
to confirm and to go back to D3/D4
to go to
|AL
0.0/20.0ppm|
to access the calibration sequences
|CAL AL L: 0.00
L 0.00: actual minimum alarm value
to stop the procedure and to go back to D5
to select the value
to confirm and to go to the maximum value insertion
|CAL AL H:20.00
H 20.00: actual maximum alarm value
to stop the procedure and to go back to D5
_
_
_
Alarm display
_
Minimum alarm value calibration
|
_
Maximum alarm value calibration
|
CL 7685
-
–
29