12. Maintenance and Cleaning
12.3.1 Replacing the fluorescent lamp
1
2
72
Fig. 73
Fig. 74
4
3
Fig. 75
Switch the instrument off and discon-
nect the power plug before replacing the
fluorescent lamp!
If the lamp is broken, it must be replaced
by the technical service, as the replace-
ment involves a high risk of injury.
• Slightly lift the sliding window (1) holding it by
the grip (2) and pull it out to the front.
Removal of the lamp
• The lamp is mounted behind a glare shield and
therefore not visible.
• Touch the lamp for better orientation.
• Lightly tilt the fluorescent tube down to the
right and pull it out of the clip (3).
• Hold the lamp with both hands and pull it to
the right out of the holder (4).
Only use lamps of the same specifica-
tion!
Installation of the new lamp
Type:
OSRAM DULUX L 18W/ 840.
• Hold the lamp in the correct mounting posi-
tion as shown and push it to the right until it
engages in the holder.
Operating Manual, Rev C, 1v3 – 04/2010