Assembling The Handwheel - Leica CM1950 Operating Manual

Uvc disinfection effective for destroying clinically relevant pathogens
Hide thumbs Also See for CM1950:
Table of Contents

Advertisement

6.
Installation
Transport with a fork lift
6.3

Assembling the handwheel

2
1
22
• The instrument can be transported with a fork lift.
To ensure a safe transport with a fork lift 3 people are
required: one operating the fork lift, and the other
2 holding the instrument on either side to prevent it
from sliding down.
• At the installation site, unscrew the adjustable feet (1) with
the open-end wrench No. 13 (see Fig. 12). This is absolutely
necessary for a stable position of the instrument.
Fig. 13, Total width with handwheel
• Insert the pin (1) of the handwheel shaft into
4
the hole (2) of the handwheel.
• Tighten the screw (4) using the No. 6 Allen key.
• Place the protective cap on the screw (4).
To dismount, proceed in reverse order.
Fig. 14
Rotate the handwheel only if the
refrigeration system is on and the
cryochamber is cold.
Operating Manual, Rev C, 1v3– 04/2010