Agilent Technologies u1273a User Manual page 149

Handheld digital multimeter
Hide thumbs Also See for u1273a:
Table of Contents

Advertisement

U1273A/U1273AX User's Guide
SAFETY COMPLIANCE
EN/IEC 61010-1:2001, ANSI/UL 61010-1:2004, and CAN/CSA-C22.2 No.
61010-1-04
MEASUREMENT CATEGORY
CAT III 1000 V/CAT IV 600 V
ELECTROMAGNETIC COMPATIBILITY (EMC)
Commercial limits compliance with EN61326-1
IP RATING
IP-54
TEMPERATURE COEFFICIENT
0.05 × (specified accuracy) / °C (from –20 °C to 18 °C, or 28 °C to 55 °C)
COMMON MODE REJECTION RATIO (CMRR)
>120 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
NORMAL MODE REJECTION RATIO (NMRR)
>60 dB at 50/60 Hz ± 0.1%
DIMENSIONS (W × H × D)
92 × 207 × 59 mm
WEIGHT
500 grams (with batteries)
WARRANTY
Please refer to
http://www.agilent.com/go/warranty_terms
• Three years for the product
• Three months for the product's standard accessories, unless otherwise
specified
• Please take note that for the product, the warranty does not cover:
Damage from contamination
Normal wear and tear of mechanical components
Manuals, fuses, and standard disposable batteries
CALIBRATION CYCLE
One year
Characteristics and Specifications
Product Characteristics
5
129

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1273ax

Table of Contents