Table 5-3 Diode specifications

Function
Diode
Notes for diode specifications:
1 Overload protection: 1000 V
2 Built-in buzzer beeps continuously when the voltage measured is less than 50 mV and beeps once for forward-biased
diode or semiconductor junctions measured between 0.3 V and 0.8 V (0.3 V ≤ reading ≤ 0.8 V).
3 Open voltage for diode: <+3.3 V
4 Open voltage for Auto-diode: <+2.5 V

Table 5-4 DC current specifications

Function
Current
Notes for DC current specifications:
1 Overload protection for 300 μA to 300 mA range: 0.44 A/1000 V; 10 × 35 mm fast-acting fuse
2 Overload protection for 3 A to 10 A range: 11 A/1000 V; 10 × 38 mm fast-acting fuse
3 Specification for 300 mA range: 440 mA continuous
4 Specification for 10 A range: 10 A continuous. Add 0.3% to the specified accuracy when measuring signals >10 A to 20 A
for 30 seconds maximum. After measuring currents >10 A, cool down the multimeter for twice the duration of the
measured time before proceeding with low current measurements.
134
Range
[3]
3 V
[4]
Auto
for short circuits with <0.3 A current
RMS
DC
and >–1.0 V
DC
DC
Range
300 μA
[1]
3000 μA
[1]
[1]
30 mA
[1][3]
300 mA
[2]
3 A
[2][4]
10 A
Resolution
0.0001 V
0.0001 V
Resolution
0.01 μA
0.1 μA
0.001 mA
0.01 mA
0.0001 A
0.001 A
Accuracy
Test current
0.5% + 5
1 mA to 2 mA
0.5% + 5
0.1 mA to 0.3 mA
Accuracy
Burden voltage
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
0.3% + 10
0.3% + 10
U1273A/U1273AX User's Guide
Approx.
Approx.
<0.04 V
<0.4 V
<0.08 V
<1.00 V
<0.1 V
<0.3 V