Preference Window for Results Reporting Options
Option
Description
REFERENCE
Reference ranges can be defined for
RANGES
each test. The ranges are shown as tic
marks on the bar graphs on the result
pages. There are no bar graphs for
blood gas, coagulation, or
immunoassay tests.
ACTION RANGES
High and low action ranges can be
defined for each test
(-99999.9–99999.9).
PRINT
Reference Ranges can be printed with
REFERENCE
results. Ranges print only if the record
RANGES
is stored with the active Preference set
in the analyzer.
OPERATOR TEST
Requires the operator to select tests to
SELECTION
be reported from a cartridge test panel.
ACT OPTIONS
The user can select between the current
(i-STAT 1 analyzer
37 °C (PREWRM) result calibration
only)
and a new NON-PREWARM (ambient
temperature) result calibration for
Celite ACT cartridges.
HEMATOCRIT
Reference anticoagulant used to
OPTIONS
calculate hematocrit result: K3EDTA
or K2EDTA/Heparin/None. (NaEDTA
is included in this option, and None
means no anticoagulant.)
CPB options:
CDS-28
Prompt: asks user whether to
apply CPB compensation when car-
tridge includes hematocrit sensor.
Never: CPB correction is never
applied when running a cartridge
with a hematocrit sensor.
Always: apply CPB correction
every time it runs a cartridge with a
hematocrit sensor.
Central Data Station
Default
Comments
– –
Ranges are displayed on the
Customization screen of the
analyzer under the Administration
Menu.
Only one range is allowed for
each test in a particular analyzer.
However, different customization
profiles can be set up in specific
analyzers used for specific patient
populations.
Make sure to enter the same
units as selected in the Unit Set
Window.
Make sure to enter Action
Disabled
Ranges within the reportable
ranges of the tests.
Make sure to enter the same
units as selected in the Unit Set
Window.
The action ranges for glucose
apply to the cartridge and the test
strip.
Disabled
The active Preference set in the
analyzer is listed as Custom on the
Analyzer Status page and the
Preference set stored with the
record is displayed on the Chart
Page when the record is recalled
and is printed with the results.
Disabled
PREWRM
K3EDTA
Analyzers can be customized by
location.
Prompt CPB