■ Delete Instrument: Click the Location/Method for the instrument(s) to be deleted and then the
serial number of the instrument to be deleted. Click Instrument > Delete... from the menu or
click Delete Inst. in the toolbar.
■ Move Instrument: Click the Location/Method for the instrument to be moved and then the
serial number of the instrument to be moved. Click Instrument > Move... from the menu, or
click Move Inst. in the toolbar. Select the new location from the drop-down list, type in a new
location, or drag and drop.
■ Find Instrument: Click Instrument > Find... from the menu, or click Find in the toolbar.
Enter the serial number of the instrument and select the Method from the drop-down list.
■ Edit Instrument Comment: Click the instrument serial number, click Instrument > Edit
Comment... from the menu or click Edit Inst. in the toolbar. Enter a comment of up to 16 char-
acters.
■ Change Current Instrument Settings
Reporting: Click the instrument serial number. Click Instrument > Current Instrument >
Change Reporting from the menu or click the down arrow next to Current Inst. and then
Reporting in the toolbar to toggle between Download and Assignment.
In Use: Click the instrument serial number. Click Instrument > Current Instrument > Toggle
In Use from the menu, or click the down arrow next to Current Inst. and then click In Use in
the toolbar to check or un-check the analyzer in the In Use column.
Note:
Instruments that do not have In Use checked do not have a download criteria
applied to them. You would use this for instruments you do not expect to be
downloaded. When an analyzer that is not marked as in use downloads, it is
set back in use and the download criteria is applied.
Monitoring: Click the instrument serial number. Click Instrument > Current Instrument >
Change Monitoring from the menu, or click the down arrow next to Current Inst. and then
Monitoring in the toolbar to check or un-check the analyzer in the Monitoring column.
■ Add Instrument Note: Click Instrument > Add Note... from the menu or click Add Note in
the toolbar. Enter a note of up to 50 characters. The note then appears as a Log Entry in the
Instrument Log tab.
■ Change Download Monitoring Criteria: Compliance to download policy can be monitored by
the CDS program. Click Instrument > Download Criteria... from the menu or click the Crite-
ria button in the toolbar. Enter the required download interval (up to 1000 hours).
Compliance with the criteria can be observed by checking the Highlight Download Status box
or by going to the Downloader Monitor. Individual Download criteria can be set for each loca-
tion and method pair.
Note:
The VetScan i-STAT 1 can also be customized either to warn the end users that
a download is required or to lockout end users if the time for a download has
been reached or exceeded. The download criteria for analyzers and for the
CDS monitor should be selected to make sense.